Live data & biofeedback
Continuous force at 80 Hz. Timer starts automatically when the pull passes the start threshold. Set a target to show a green zone on the graph — the patient holds the line (isometric biofeedback / graded loading).
How to do it
- Tare with nothing pulling. Press Start.
- Patient pulls/pushes. The timer starts by itself once force passes the threshold.
- For biofeedback: enter a target (e.g. 60% of their MVC) — the patient keeps the line inside the green zone.
- Press Stop when done. Average = mean force while above threshold.
Peak force (MVC – maximal voluntary contraction)
The main MMT number. Best of several maximal efforts per side. Gives peak kg/N, best 1-second average, % body weight, torque (if lever arm entered) and left–right symmetry.
How to do it (e.g. knee extension)
- Patient seated, knee at 60°, ankle cuff 2 cm above the ankle, strap to plinth leg, strap at 90° to the shin, no slack.
- Optional: measure lever arm = knee joint line to strap centre (cm) for torque.
- Press Start. On PULL: “push as hard as you can – push, push, push” for 5 s. Rest during REST.
- Best two trials should be within 10%. If not, run one more trial.
- Symmetry = weaker ÷ stronger × 100. ≥ 90% is a common return-to-sport level.
Rate of force development (RFD) – explosive strength
How fast force rises. Patient pulls as hard and as fast as possible. Reported two standard ways: 20–80% of peak, and 0–X ms from onset.
How to do it
- Same set-up as peak force. Patient relaxed — strap just taut, force near zero before the cue (no pre-tension).
- On PULL: “as fast and hard as you can” — a quick explosive push, hold ~2 s.
- Trials with a counter-movement (force dipping below zero first) or pre-tension are flagged.
- Useful for sports, ACL return-to-sport and falls-risk in elderly.
Endurance – hold a target as long as possible
Patient holds a set % of their MVC. Ends automatically when force drops below the zone for more than 1 s. Gives time to failure, time in zone and force drift.
How to do it
- Run Peak force first — MVC fills in automatically (or type it).
- Choose the target, e.g. 50% MVC. Press Start.
- Patient builds up to the green zone and holds it. Test ends when they fall below the zone for >1 s, or at max time.
- Use for tendinopathy load tolerance, postural/trunk endurance, and fatigue tracking.
Repeaters – interval work / rest
Timed sets of contractions with rest, with an optional target zone. Gives per-rep force, time in zone and fatigue across reps. Used for training and isometric rehab programmes.
How to do it
- Optional: run Peak force first and set a target (e.g. 70% MVC). Without a target, every rep is maximal.
- Press Start. Patient pulls on PULL, relaxes on REST. Beeps mark each change.
- Fatigue = last-rep average ÷ first-rep average.
Critical force (CF) – 4-minute all-out test
24 maximal efforts: 7 s pull, 3 s rest. Force falls and levels off. CF = average of the last 6 reps (the force the muscle can sustain); W′ = work done above CF (reserve capacity). Advanced / sports test.
How to do it
- Warm up. Patient should be fresh (no hard training that day).
- Press Start. Every PULL is maximal — do not pace. Relax completely on REST.
- Very tiring. Not for acute injuries, post-op early stages, or cardiac-risk patients.
- Originally validated on finger flexors (climbing); use for other muscles as a tracking measure.
Accuracy check (known weight)
Verifies the device. Hang a known weight and compare. Device spec: ±1.25 kg (0.25% of 500 kg).
How to do it
- Hang the Progressor from a fixed point. Press Tare with nothing attached.
- Hang a weight you know exactly (gym plate, or a bag weighed on a shop scale). Let it stop swinging.
- Enter the weight and press Start. Repeat with 2–3 different weights.
JSON data
Device
Patient & settings
Session results
Raw Bluetooth log
Opening this page
- Chrome on a laptop or Android tablet (not iPhone/iPad). Bluetooth on (and Location on Android).
- Press the Progressor’s button to switch it on, then Connect Progressor.
- On Linux, if Bluetooth is “not available”, enable
chrome://flags/#enable-experimental-web-platform-features. - No device? Simulator sends fake data in the exact Tindeq format.